Show simple item record

dc.contributor.authorYaras, Y. S.
dc.contributor.authorCakmak, O.
dc.contributor.authorGunduz, A. B.
dc.contributor.authorSaglam, G.
dc.contributor.authorOlcer, S.
dc.contributor.authorMostafazadeh, A.
dc.contributor.authorBaris, I.
dc.contributor.authorCivitci, F.
dc.contributor.authorYaralıoğlu, Göksen Göksenin
dc.contributor.authorUrey, H.
dc.date.accessioned2017-06-17T13:54:22Z
dc.date.available2017-06-17T13:54:22Z
dc.date.issued2017
dc.identifier.issn0277-786X
dc.identifier.urihttp://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=2608614
dc.identifier.urihttp://hdl.handle.net/10679/5361
dc.description.abstractWe developed two types of cantilever-based biosensors for portable diagnostics applications. One sensor is based on MEMS cantilever chip mounted in a microfluidic channel and the other sensor is based on a movable optical fiber placed across a microfluidic channel. Both types of sensors were aimed at direct mechanical measurement of coagulation time in a disposable cartridge using plasma or whole blood samples. There are several similarities and also some important differences between the MEMS based and the optical fiber based solutions. The aim of this paper is to provide a comparison between the two solutions and the results. For both types of sensors, actuation of the cantilever or the moving fiber is achieved using an electro coil and the readout is optical. Since both the actuation and sensing are remote, no electrical connections are required for the cartridge. Therefore it is possible to build low cost disposable cartridges. The reader unit for the cartridge contains light sources, photodetectors, the electro coil, a heater, analog electronics, and a microprocessor. The reader unit has different optical interfaces for the cartridges that have MEMS cantilevers and moving fibers. MEMS based platform has better sensitivity but optomechanical alignment is a challenge and measurements with whole blood were not possible due to high scattering of light by the red blood cells. Fiber sensor based platform has relaxed optomechanical tolerances, ease of manufacturing, and it allows measurements in whole blood. Both sensors were tested using control plasma samples for activated-Partial-Thromboplastin-Time (aPTT) measurements. Control plasma test results matched with the manufacturer’s datasheet. Optical fiber based system was tested for aPTT tests with human whole blood samples and the proposed platform provided repeatable test results making the system method of choice for portable diagnostics.
dc.description.sponsorshipTÜBİTAK
dc.language.isoengen_US
dc.publisherSPIE en_US
dc.relationinfo:turkey/grantAgreement/TUBITAK/113S074
dc.relation.ispartofProc. SPIE 10055, Optics and Biophotonics in Low-Resource Settings III
dc.rightsrestrictedAccess
dc.titleDisposable cartridge biosensor platform for portable diagnosticsen_US
dc.typeConference paperen_US
dc.peerreviewedyes
dc.publicationstatuspublisheden_US
dc.contributor.departmentÖzyeğin University
dc.contributor.authorID(ORCID 0000-0002-3242-5797 & YÖK ID 192045) Yaralıoğlu, Göksenin
dc.contributor.ozuauthorYaralıoğlu, Göksen Göksenin
dc.identifier.wosWOS:000401131300019
dc.subject.keywordsCoagulation
dc.subject.keywordsBiosensor
dc.subject.keywordsOptical read out
dc.subject.keywordsMEMS
dc.subject.keywordsFiber sensor
dc.subject.keywordsaPTT
dc.subject.keywordsPoint of care
dc.contributor.authorMale1
dc.relation.publicationcategoryConference Paper - International - Institutional Academic Staff


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record


Share this page